Joan Laboratory P20 P200 Micropipette Autoclavable Technical Data | ||||||||
Picture | Model | Volume range |
Increment (ul) |
Test volume (ul) |
Max permissible systematic error (Inaccuracy) |
Max permissible random error (Imprecision) |
||
% | ul | % | ul | |||||
![]() |
M8-10 | 0.5 - 10ul | 0.1 |
10 5 1 |
1.50 2.50 4.00 |
0.15 0.125 0.04 |
1.50 2.50 4.00 |
0.15 0.125 0.04 |
M8-50 | 5-50ul | 0.5 |
50 25 5 |
0.50 1.00 2.00 |
0.50 0.375 0.15 |
0.50 1.00 2.00 |
0.25 0.25 0.10 |
|
M8-300 | 50-300ul | 5 |
300 150 50 |
0.25 0.50 0.80 |
2.1 1.5 0.75 |
0.25 0.50 0.80 |
0.75 0.75 0.4 |
|
![]() |
M12-10 | 0.5-10UL | 0.1 |
10 5 1 |
1.50 2.50 4.00 |
0.15 0.125 0.04 |
1.50 2.50 4.00 |
0.15 0.125 0.04 |
M12-50 | 5-50ul | 0.5 |
50 25 5 |
0.50 1.00 2.00 |
0.5 0.375 0.15 |
0.50 1.00 2.00 |
0.25 0.25 0.10 |
|
M12-300 | 50-300ul | 5 |
300 150 50 |
0.25 0.50 0.50 |
2.1 1.5 0.75 |
0.25 0.50 0.80 |
0.75 0.75 0.40 |
|
Biohit tips can also be used in our product! |